Wireless Affective Devices

From AIRWiki
Revision as of 15:50, 25 July 2009 by PaoloVilla (Talk | contribs)

Jump to: navigation, search

Part 1: project profile

Project name

Wireless Affective Devices (before 2008/10/30 only Affective Device)

Project short description

This project aims the development of devices for biological signals acquisition that can be used for Affective computing applications.

Dates

Start date: 2008/01/08

Preliminary release end: 2008/08/08

End date: Unknown (WIP)

Internet site(s)

People involved

Project heads

Other Politecnico di Milano people

Students currently working on the project

Students that have worked on the project

  • Mattia Colombo (Blood Volume Pressure sensor)
  • Alessia Cornaggia (Galvanic skin response sensor)
  • Sala Mirko (Galvanic skin response sensor)

Laboratory work and risk analysis

Laboratory work for this project will be mainly performed at AIRLab/Lambrate. It will include electrical and electronic activity. Potentially risky activities are the following:

  • Use of mechanical tools. Standard safety measures described in Safety norms will be followed.
  • Use of soldering iron. Standard safety measures described in Safety norms will be followed.
  • Use of high-voltage circuits. Special gloves and a current limiter will be used. Standard safety measures described in Safety norms will be followed.


Part 2: introduction (why we want do this)

Everyone knows what an emotion is, until asked to give a definition”
                                        
                                                                Beverly Fehr and James Russell

Research on Human-Computer-Interfaces has recently begun to take into account the affective state of the user. By appropriately reacting to the affective user state, interfaces could not only become more pleasant or entertaining, but also more effective or safer. It is known that affective states have bodily correlates, and thus physiological signals could provide the necessary information for acquiring the affective user state. In contrast to other sources of information about the affective user state such as speech or facial expression, most physiological signals are not under voluntary control, and thus cannot be masked up to the same extent. Furthermore, it can be presumed that the signs of affective states in physiological signals are less dependent on individual and contextual factors: this is because the signals correspond to internal physiology, largely related to the autonomous nervous and limbic systems, rather than to external expressions that can be manipulated easily. Recently, the issue of emotion regulation drew interests from several fields including neuroscience, psychology, education, computer science, etc. Possible applications are e-learningin, human computer interaction, development of communicative technologies for use by people with autism, car infotainment system, lie detection tests, cognitive workload, multimodal user interfaces MMUI (allow users to control computers using speech and gesture), etc.


Part 3: project description

The project is aimed at developing a device, connected to a PC, to acquire biological signals (such as GSR, BVP, Temperature and Head orientation) in a way that is less invasive for the subject. The term invasive is related to how much the sensors we use compromise the normal interaction between subject and machine. Through these biological signal the PC is able o understand emotion in people who wear the device.

An affective device should be weared easily, should not have cables and should acquire as much sensors as possible. We relax some constraint on the quality of signal but we add constraint on way they are acquired(i.e. the subject can put it on easily, the usual movements are not influenced..). The aim of the resulting device is not to replace the standard medical devices usually used to acquire that signals, but is to be much more suitable for industrial application in which the medical requirement are replaced by invasivity requirement.

It is composed by:

  • Low power, high efficiency, battery powered design
  • Wireless comunication with a PC
  • Blood Volume Pressure Sensor (Preliminary release by Mattia Colombo)
  • Galvanic Skin Response Sensor (Preliminary release by Sala Mirko e Alessia Cornaggia)
  • Temperature Sensor
  • 3 Way Accelerometer
  • Possibility of further extension


GSR (or EDR or EDA)

An often misunderstood and difficult technique, GSR (also known as the electrodermal response EDR or electrodermal activity EDA) has gone through many phases of interest and rejection since the early 1900's and is very popular in psychophysiological studies since Carl Jung and his students (1907) described it as a mean to enter the “sea of the unconscious” because “every stimulus accompanied by an emotion produced a deviation of the galvanometer” directly proportional to the strength of the emotion aroused. It has been used in important research on anxiety and stress levels (Fenz & Epstein, ' 67); and it has been a part of lie detection (Raskin, ' 73). Controversy has centered around the technique, underlying mechanisms, and the meaning of the responses obtained from the skin. There has been a long history of electrodermal activity research, even if most investigators accept the phenomenon without understanding exactly what it means (Hume, ' 76). The resistance of the skin is usually large, approximately 1M; however, momentary changes in the level of the sweat gland activity causes changes in resistance (up to approximately 950K). Physiology, the GSR reflescts sweat gland activity and changes in the sympathetic nervous system and measurement variables. The activity of the sweat glands in response to sympathetic nervous stimulation ( Increased sympathetic activation ) results in an increase in the level of conductance. There ia a relationship between sympathetic activity and emotional arousal, although one cannot identify the specific emotion being elicited. Basically there are two techniques in the history of Electrodermal measurement. In one a current is passed thru the skin and the the resistance to passage is measured; in the other no current is used externally and the skin itself is the source of electrical activity. By applying a conventional 0.5 Volts across the skin and measuring changes in the corresponding conductance, the emotional state of the subject can be inferred. In general, the electrodes used are of the Ag/AgCl type which are recessed from the skin and require the use of a suitable electrode paste. It is important to note, however, that fluctuations in skin conductivity are resultant of many types of arousal. By observing only these changes, it is impossible to deduce without prior knowledge whether the subject has become happy, startled, physically active, etcetera. EDA consists of two components: tonic and phasic. The tonic component is a low frequency baseline conductivity level, which can oscillate over the course of days. The phasic component rides on top of the tonic component, exhibits more rapid fluctuations, and generally increases when a person is aroused. Problematically, each person has a different tonic conductivity, so in order to infer the arousal level of the subject, the relative changes in EDA must be analyzed over a period of time. Furthermore, skin conductance (measured in units of siemens; formerly mhos) depends on the skin path length between the two electrodes contacts, even for subjects with identical skin conductivity (measured in units of siemens/meter). It is for these reasons that it is crucial to analyze the temporal variations of the EDA signal.

BVP

The cardiovascular system is in charge to keep us alive by maintaining the vital blood flow throughout the body, providing nutrients and oxygen to our cells. The SNS controls this system by shifting the flow in response to exercise, temperature,postural and emotional changes. Blood Volume Pulse (BVP) is the most common measure of vasomotor activity because it reflects the phasic pumping of the heart, the vasodilation of vessels that changes in the amount of acral blood delivered. The Blood Volume Pulse (BVP) waveform is an important indicator of circulatory function that can be obtained noninvasively through a photoplethysmographic transducer (a PPG is often obtained by using a pulse oximeter, Shelley and Shelley, 2001) applied to the finger or the earlobe of human subjects. In particular, this signal may be useful in determining the degree of cardiovascular change undergone by a subject through an exercise session.

Principles of photoplethysmographic Technology:

The principle of photoplethysmographic (That is the same of pulse oximetry) is based on the red and infrared light absorption characteristics of oxygenated and deoxygenated hemoglobin. Oxygenated hemoglobin absorbs more infrared light and allows more red light to pass through. Deoxygenated (or reduced) hemoglobin absorbs more red light and allows more infrared light to pass through. Red light is in the 600-750 nm wavelength light band. Pulse oximetry uses a light emitter with red and infrared LEDs that shines through a reasonably translucent site with good blood flow. Opposite the emitter is a photodetector that receives the light that passes through the measuring site. There are two methods of sending light through the measuring site: transmission and reflectance. In the transmission the emitter and photodetector are opposite of each other with the measuring site in-between. The light can then pass through the site. In the reflectance method, the emitter and photodetector are next to each other on top the measuring site. The light bounces from the emitter to the detector across the site. The transmission method is the most common type used. While pulse oximeters are a commonly used medical device the PPG derived from them is rarely displayed, and is nominally only processed to determine heart rate.

Temperature

The temperatures of peripheral limbs vary as a consequence of changes in blood flow.


Previous works

Here is a preliminary release of the headset produced before 2008/08/08

Wd1.jpg

If you are interested on the making of the headset please see making rubber models

* Blood Volume pressure First graph showing a signal acquired with the wireless sensor. Bvp.png

* Skin galvanic response (Sala Mirko e Alessia Cornaggia[1]) Galvanic skin response (GSR), also known as electrodermal response (EDR), psychogalvanic reflex (PGR), or skin conductance response (SCR), is a method of measuring the electrical resistance of the skin. There is a relationship between sympathetic activity and emotional arousal, although one cannot identify the specific emotion being elicited. The GSR is highly sensitive to emotions in some people. Fear, anger, startle response, orienting response and sexual feelings are all among the emotions which may produce similar GSR responses. Response of the skin to the passage of a small electric current. The ease with which the current flows between two points on the skin can be used to indicate stress. When a person is tense or emotional, the sweat glands become more active, increasing moisture on the skin; this allows the electric current to flow more readily. The response may also be used in relaxation training: information about the galvanic skin response is fed back aurally or visually to the subject who can, with practice, learn to increase or decrease sweating on the skin by learning to relax or tense muscles. We measure the galvanic skin response using two electrodes attached to the skin. The optimal placement is the palm of the hands, but we are experimenting other places like on the forehead. The circuit we are creating is the one below.

Circuito.jpg


Part 4: project the device

Re-definition of the complete work

Sensors are an important part of an Affective Computing System because they provide information about the wearer's physical state or behavior. They can gather data in a continuous way without having to interrupt the user. To define in detail electronic specifications and requirements it's necessary a preliminary research/study (in literature and web) about state of art of sensor type used to monitor each type of biological signals we are interested about. In this way it's possible both understand problematics related to interaction between biological and electonical worlds and improve or re-desig existing prototype.

About Sensors:

BVP will be acquire by an IR transmitter-receiver measuring light reflected from skin (photoplethysmography). Particular attention will be put on position, geometry and power absorption of this stage.

GSR will be acquiret by a couple of electrodes forcing a low current (or voltage) and reading voltage (or current) through electrodes. Particular attention will be put on electrodes position and material. In addition is possible to foresee some problems in amplifier stage due to large range of skin resistance. Problem consist in chosing a correct amplification to overwork ADC full scale with a signal affected by little variation but huge possible range.

Temperature will be acquired by thermocouple or PTC/NTC. Slowly variable signal, possible problems with noise due to resolution required.

Head orientation will be acquired by an accelerometer and a gyroscope, problably integrated in the same chip. Two possibilities: analog output (need to digitalize up to six line) or digital output (like displaied in schematic block diagram).


This was the preliminary simplified schematic block diagram of the complete instrumentation setup:

WADSchematicDiagram01.jpg

NOTE:

* In schematic is not present Power Supply Module: it will be based on a rechargeable Li-ion
battery. It must satisfy requirements of high efficiency, logn life and low weight.

* ADC and Microcontroller could be unified using a microcontroller with internal ADC and
internal or external multiplexer to select lines.

* Wireless section could be implemeted in Bluetooth (instead of using a XBee module like in
preliminary release) due to it's diffusion in cellular phones, notebooks, MP3 reader...
without necessity of external modules.

It was based on a theoretical approach to any single part. Any sensing device was designed in "conventional" way founded on analog electronic (a bridge with a NTC resisistance as sensing element, discrete IR led and some photodiode around to capture the most reflefted light as possible, GSR electrodes on the gain arm of a non-inverting amplifier...). Unfortunatly a deep analysis on signal physiology, signal transduction, noise vs required accurancy, forced us to a completely different approach implementing everywhere possible a digital interface.

Sensors Design

BVP

Discrete IR leds and phototransistor, used in the first realize (see fig.), had been replaced first with a single led and one or more photodiodes to optimize power consumption in case of non-planar sensor positioning. Subsequently an integrated Reflective Object Sensor was introduced, with benefits about space saving, transmitter-receiver matching and power consumption. To ensure light power stability, led is powered through a linear led drive LDO (Low Drop-Out). The choise of linear LDO is necessary to ensure no noise or spikes would be generated by charge pump systems or boost converters. In addition, the linear drive utilized requires only one external resitor to set led current and a decoupling capacitor, reducing part counts and board space occupation. The couple factor of the reflex sensors is relatively small, even in the case of good reflecting surfaces. Therefore an additional amplifier is necessary at the sensor output: classic trans-impedance amplifier used with photo-diodes and photo-transistors was introduced. This circuit features an output voltage that is a linear function of the light level on the photo-transistor. The gain is set by the feedback resistor which could be in the millions of ohms. Linear operation is achieved by operating the photo device with a fixed voltage drop across it. Trans-impedance amplifier is followed by an active fourth order band-pass filter (0.5Hz-5Hz): it's composed cascading two second order bandpass cells. In first project session only a single cell was used, but a deeper study combined with previous experiences suggests to use a 4th order filter to reduce 50Hz interference (50Hz interference produced by mains is placed only a decade after the filter pole). Filter is followed by a VGA (Variable Gain Amplifier) digitally controlled through I2C Bus that allows to optimize dinamically ADC input signal amplitude directly during acquisition session, obtaining the best performance from ADC. The resulting circuit is quite compact (led drive, reflective sensor, trans-impedence amplifier, active filter and variable-gain amplifier), doesn't require much power and represents a good mix between analog and digital approach: analog front-end with acquisition and first conditioning, and at the chain-end a digital controlled amplifier to ensure flexibility.

WADBVPComparison.jpg

GSR

GSR will be acquired by a couple of electrodes forcing a low voltage and reading current through electrodes. First designs and implementations were based on literature study: best solution appears to put GSR electrodes on the gain arm of a non-inverting amplifier (see fig.) rather than using a Wheatstone bridge due to week signal produced by this solution. Putting GSR electrodes on the gain arm produces a sort of gain modulation of the constant voltage applied between electrodes but doesn't solve problems in amplifier stage due to large range of skin resistance. Problems in chosing correct amplification to overwork ADC full scale with a signal affected by little variation but huge possible range could be reduced usign a variable resistor on gain arm (for example a digital potentiometer controlled by the microcontroller), but cannot be completely removed. In addition little signals are easily deteriorated by noise. Solution is provided considering a new point of view: measure of unknown skin resistence is extremely similar to measure unknown electrical impedence with an auto-range multimeter! Solution consist in a digital Impedance Converter: AD5934 is a high precision impedance converter system which combines an on-board frequency generator with a 12-bit, 250 kSPS, analog-to-digital converter (ADC). The frequency generator allows an external complex impedance to be excited with a known frequency; the response signal from the impedance is sampled by the on-board ADC and a discrete Fourier transform (DFT) is processed by an on-board DSP engine. The DFT algorithm returns a real (R) and imaginary (I) data-word at each output frequency. This solution not only allows to measure impedence range from 100 Ohm to 10 MOhm, but in adition allow to investigate it across a wide range of frequency, producing a noise-free digital output. The resulting circuit is extremely compact (converter chip, one setting resistor and bypass capacitors), provides a system accuracy of 0.5% and connects to microcontroller using I2C interface, ensuring an high flexibility and the possibility to modify acquisition settings after board production (through microcontroller firmware).

WADGSRComparison.jpg

Temperature

First idea was to acquire temperature by a NTC connected in a Wheatstone bridge. Trying to reduce noise, bridge was powered by an oscillator circuit with a frequency of some KHerz, to translate signal out of 1/f noise region. Bridge outputs were connected to an INA amplifier wich also implements signal filtering (see fig.). Weak signal, noise and resolution required (0.1°C) put in crisis the system. In addition other unsolved problems were the optimization of thermal and mechanical contact between sensor and body and the sensor large thermic constant. Unstable thermic contact produces unreliable measures, whereas high sensor thermic constant hampers a fast sensor response (needed to detect short peak due to emotional reactions). Possible solutions to problems of weak signal and noise arrive from digital integrated temperature sensor (for example LM35 or LM74), but these don't solve the problem of thermal contact and thermal constant. In addition none of the sensors take in consideration provides the required resolution. High resolution required, noise immunity, fast response and thermal contact immunity problems had been solved using a digital optical infra red thermometer. The MLX90614 is an infra red thermometer for non contact temperature measurements based on a thermopile detector chip including signal conditioning in the same TO-39 package. Thanks to its low noise amplifier, 17-bit ADC and powerful DSP unit, a high accuracy and resolution of the thermometer is achieved. The calculated object and ambient temperatures are available in RAM of MLX90614 with 0.02°C resolution: they are accessible by 2 wire serial SMBus compatible protocol. A special version for medical applications offers an accuracy of ±0.1ºC in a limited temperature range around the human body temperature. Small size, no external component requirements and power saving mode allow us to obtain a perfect matching between required performances and space saving.

WADTemperatureComparison.jpg

Accelerometer

Head orientation and movements will be acquired by an accelerometer. First we think to utilize an ST Microelectronic LIS3L02AS4, due to the availability in the Lab of the evaluation board STEVAL-MKI001. It is a 3-axis ±2g/±6g linear accelerometer with 3 analog outputs, so it requires an external ADC to digitalize the three axis acceleration values. The conversion could be performed by a microcontroller with internal ADC. A fast Internet research reveled this sensor is now obsolete and out of production. The choise had been to substitute the old chip with the new one: the LIS302DL (still produced by ST Microelectronic). It's a 3-axis, ± 2g/± 8g smart digital output accelerometer with I2C/SPI digital output interface and really small package. The device is capable of measuring accelerations with an output data rate of 100 Hz or 400 Hz and the self-test capability allows the user to check the functioning of the sensor in the final application. In addition it may be configured to generate inertial wake-up/free-fall interrupt signals when a programmable acceleration threshold is crossed at least in one of the three axes. Thresholds and timing of interrupt generators are programmable by the end user on the fly. This device, providing a digital output, doesn't need signals conversion, guaranteing better noise immunity; programmable interrupts and test interface allow greater fexibility in project. In addition wery small package, only one external resistor requirement (plus baypass capacitor) reduce significantly board space requirements.

WADAccelerometer.jpg


NOTE:

* In previous schematics, components phisical dimensions are not in scale.


Microcontroller

Foundamental requirements:

- High speed to ensure high data output rate even in presence of a large number af sensing elemets;

- Fast Internal ADC with multiple inputs;

- (At Least) One I2C interface;

- (At Least) One SPI interface;

- (At Least) One UART interface;

- Low-power consumption;

- Power supply 3.3V;

- Small package;

In addition microcontroller was chosen between Microchip PIC family devices, due to positive previous experiences and availability of software and programmer in the Lab.


PIC24HJ256GP206 is a 16 bit microcontroller capable of 40 MIPS speed. It provides 256KB of Flash memory, 16KB of Ram, a 10-bit 1.1Msps or 12-bit 500Ksps ADC, UART, SPI and I2C interface in a 64 pin TQFP package (10mmX10mm) powerd between 3.0V and 3.6V. Power core is provided by an internal regulator reducing parts count on the board.

The device is able to provide at the same time one I2C interface, two SPI interface and one UART interface: another UART port shares pins with an SPI port and a second I2C port shares pins with the second UART interface. Findig a device that provides at the same time, on indipendent pins, an I2C, a SPI and an UART interface reduces considerably possible choises and suggests to chose a high performance microcontroller. At the same time, high performance solutions have a larger pin counts than standard microcontroller, but this characteristic can be used to ensure further expansibility to the system. The second SPI interface can be used to connect another "intelligent" device or a mass storage device, for example a SD (Secure Digital) card to storage acquired data (see fig.).

WADSDCard.jpg

Input pins connected to ADC and not used at the moment could be collected to a connector for further extension modules (for example an EMG or ECG peripheral or any other analog signal interesting in a particolar application). For the same reason any other pin non used at the current stage is reported to an extension connector. To guarante maximum flexibility an further update, board provides both ICD and JTAG connections (allowing on-board programming and debugging) through a couple of indipendent dedicated connectors fully compatibles with Microchip programmers and evaluation boards.

WADMicro.jpg

Wireless Connection

Wireless connection is based on BlueGiga WT12 module: it's a next-generation, class 2, Bluetooth 2.0+EDR (Enhanced Data Rates) module containing all the necessary elements from Bluetooth radio to antenna and a fully implemented protocol stack. By default WT12 module is equipped with powerful and easy-to-use iWRAP firmware that enables users to access Bluetooth functionality with simple ASCII commands delivered to the module over serial interface like a Bluetooth modem. The dedicated onboard microcontroller (MCU) acts as interrupt controller and event timer run the Bluetooth software stack and control the radio and host interfaces. A 16-bit reduced instruction set computer (RISC) microcontroller is used for low power consumption and efficient use of memory. WT12 module support USB, SPI and UART interface and UART interface is used in this project to connect PIC microcontroller to wireless transmitter (see fig.).

WADWireless.jpg

One of the most important WT12 characteristic is the high integration between all parts needed for efficent Bluetooth transmission: user interface, controller, radio circuits and antenna, all included in a low power, low size board (25.6mmx15mm) that requires a minimum number of external components. Many other bluetooth components ara available, such as LMX9830, produced by National Semiconductor. It's an extremely small device (only 6.1mm×9.1 mm), but requires a number of external components (see fig.) like oscillators, resistors, capacitors, inductors and the antenna, in addition to an high competence in radio frequency layout circuits.

WADLMX9830.jpg

Lack of experience in radio frequency circuit suggests to chose a ready-to-use module, factory optimized for best performance and easy to integrate in our board.


This is the updated simplified schematic block diagram:

WADSchematicDiagram02.jpg


Part 5: realize a prototype [WIP (Work in Progress)]

Fist stage plans to realize a board using, where possible, through-hole (TH) compenents soldered on a prototyping board (100x160mm) powered from a laboratory power supply (instead of Li-on battery) and not implementing bluetooth wireless transmission. Parts not available on TH package are used with special SMD->TH adapter boards or simply fixing and soldering them to the PCB by wires.

Second stage plans to realize microcontoller firmware and test it acquiring biological signals from a patient. Firmware will be developed by subsequently steps, implementing and testing only a single peripheral at each time: only at the end single pheripheral softwares will be merged together.

Third stage will be bluetooth interface (hardware and software) implementation and integration on the board with transmission test.

Fourth stage will consider power supply optimization with battery management and recharge interface.

Finally, only after a successful full board test, the last stage will be the circuit layout design with an electronic CAD, using where possible, Surface Mount Devices (SMD) to reduce board area.


Main board

Main board include power supply regulator with led power-on indicator, PIC microcontroller (mounted on a SMD->TH adapeter board) with basic circuitry (decoupling capacitors, oscillator and programming-debugging ICD interface), a status led connected to PIC to verify basic microcontroller function, a series of connectors for add-on modules (BVP, GSR, temperature and accelerometer) and a USB PC interface not considered before.


USB interface

USB interface, not projected before, replaces in developing stage Bluetooth interface, allowing a direct control/communication with a PC. In this prototype PIC UART, althought being connected to Bluettoth module, is connected to a FTDI FT232RL chip. The FT232RL is the latest device to be added to FTDI’s range of USB UART interface Integrated Circuit Devices. USB to serial designs using the FT232RL have been further simplified by fully integrating the external EEPROM, clock circuit and USB resistors onto the device. The FT232RL is available in Pb-free (RoHS compliant) compact 28-Lead SSOP, so this require another SMD->TH board adapter but the circuit design is very simple, including two status led for TX and RX operations. An USB B female connector was eputated the best choise for PC connection allowing to use a simple USB printer cable.

WADUSBInterface.jpg


Note that USB and Bluetooth must not be connected to the board at the same time! USB port has only a developent function.


BVP

Circuit designed for BVP acquisition (refer to previous schematic) doesn't meet the expected results. Week signal hanged by noise suggested us to re-design the circuit simplifying it and reducing the number of components. For this design also sensing element has been substitute. Instead of an integrated reflective IR sensor (SFH9202), new project use a couple af discrete IR led and photothansistor matched: SFH487 and SFH309FA. SFH487 is a very highly efficient GaAlAs infrared emitter (880 nm) and SFH309FA is a silicon NPN phototransistor especially suitable for applications of 880 nm with an high linearity. Both have a 3 mm led plastic package, are spectrum-matched and receiver is ambient-light shilded. After a series of positive test without the led driver (not used due to the really reduced mechanical dimensions), also led drive circuit was removed and replaced with a resistance connected directly to a PIC pin. This choise reduces parts count and board space required without making worse sistem performance, allowing (like led driver) to turn off the sensor led when not used to reduce power consumption.

WADBVPNew.jpg

Even if operation amplifier used are half of before, double second-order filter stage is manteined. In addition to filtering, these opamp privide a first gain (note that in previous design filtering stages works as a unitary buffer in pass-band) allowing a high signal at VGA (variable gain amplifier) input stage allowing a better SNR. 1N4148 diode provides a sort of ofsetted virtual ground, shifting up all signals but not being amplified through the amplification chain. Figure show the signal measured directly on filtering stage output (TS952 pin 7) before the VGA with a Tektronix scopemeter TDS2014B. No particular triks had been used: signal has been recorded the first time circuit was assembled on a protoboard.

WADBVPScope03.jpg


GSR

First approach with AD5934 resulted really hard due to some errors occurred in control software in relation with internal chip complexity and I2C bus protocol. After a sart session with the AD5934, GSR hardware acquisition interface was further simplified replacing it with AD5933 (refer to schematic). AD5933 has the same pinout and functions of AD5934 but provides also an internal oscillator at 16.776MHz allowing no external crystal connection. A temperature sensor that allow the user to measure chip temperature and 1 MSPS Converter (instad of a 250KSPS one) are the only other differences between the tho integrated semiconductor produced by Analog Devices. On the other hand AD5934 has a larger impedance measurement range that goes from 100 Ohm to 10 MOhm in opposition to AD5933 that can only measure impedences between 1 Ohm and 10 MOhm. Because of the high skin impedence values (in the range of MOhm for dry fingers skin), this is not a problem in our application.

WADGSRNew.jpg

First dataset acquired whith the correct control software and the correct data reading routine, has shown all the potentiality of this chip, providing data comparables with RAW data acquired with a bigger and very expensive professional equipement produced by ProComp.


ACCELERATION

Lis302DL datasheet gives a short definition of this chip as "MEMS motion sensor 3-axis - ± 2g/± 8g smart digital output “piccolo” accelerometer" and it is really very, very small for human-working. For this space saving quality this accelerometer is also use in third generation Apple i-phone. The major problem with this sensor had been to solder it to a sort of adapter to connect it to the circuit, but after that it works fine, with a good sensitivity and accuracy. Because of free fall and double click recognition at not required (at the moment) in this application, software routines for setting configuration registers and reading data resulted simplified. A usefull caracteristic of this sensor, not previously noticed, is the presence of internal pull-up resistors that allows no necessity of external I2C bus pull-up resistors, reducing space occupation and components count.


TEMPERATURE

The choice to use MLX90614 due to his characteristics (non contact measure, 0.02°C resolution with accuracy of ±0.1°C for medical applications version) creates a really serious supplies problem: Melexis (sensor manufacture) wasn't able to give us a sample and no one in Italy can sell us only a pair of theese sensors to try them. A single piece was bought from Future Electronics, ordering it via web and paying it via credit card. Fortunately the sensor, that cames directly from USA, arrived in Italy few days later the payment in excellent conditions.

WADMLX90614.jpg

Problems occurred interfacing this sensor on I2C bus with other devices (AD5933 and LIS302) connected. MLX90614 has a 2 wire serial SMBus compatible protocol: I2C bus and SMBus are popular 2-wire buses that are essentially compatible with each other, but are not the same. The buses operate at the same speed, up to 100kHz, but the I2C bus has both 400kHz and 2MHz versions. Complete compatibility between both buses is ensured only below 100kHz. Timeout and (as a consequence of timeout) minimum clock speed are the most important differences between the I2C bus and the SMBus:

I2C Bus = DC (no timeout) SMBus = 10kHz (35mS timeout)

Despite differences in logic-level specifications between the two buses, generally devices can be mixed and matched with abandon. Level Specifications for the I2C Bus and the SMBus are shown below (VDD is typically expected to be between 3 volts and 5 volts):

High I2C VDD Dependent 0.7 × VDD

I2C Fixed 3.0V

SMBus 2.1V

Low I2C VDD Dependent 0.3 × VDD

I2C Fixed 1.5V

SMBus 0.8V

What wasn't readily evident in the discussion of voltage levels is that these levels are also specified with different currents between the buses. The SMBus specifies a minimum sink current of 100microA, and a maximum of 350microA, compared to 3mA for the I2C bus. This in turn would determine the lowest acceptable value of the pullup resistor, examples of which are shown in the table below.

3V VDD 5V VDD

I2C Bus > 1k > 1.6k

SMBus > 8.5k > 14k

In reality, it is not unusual to encounter pullup resistor values in SMBus systems, which violate this specification by being lower than recommended. A very popular range for pullup resistor values, even in some SMBus systems, seems to be 2.4k to 3.9k.